Model
|
Turb PLUS 2020
|
Turb PLUS 2120
|
Measurement principle
|
Scattered light measurement white light, nephelometric measurement
|
Scattered light measurement infrared light, nephelometric measurement
|
Measurement Range
|
Factory setting: 0 ... 100 NTU; optional measuring ranges 0 ... 10 and 0 ... 1000 NTU (loaded on USB)
|
Resolution
|
Down to 0.0001 NTU
|
Accuracy
|
±2 % of reading or ±0.02 NTU below 40 NTU whichever is greater (0–100 and 0–1000 ranges) ±5 % of reading above 40 NTU (0–100 and 0–1000 NTU ranges) ±2 % of reading or ±0.02 NTU (0–10 NTU range)
|
Limit of Detection
|
0.02 mg/l
|
Response time
|
Adjustable (1 ... 60 readings)
|
Sample Temperature
|
1 °C ... 50 °C (34 °F ... 122 °F)
|
Sample Flow Rate
|
100 ml/min. ... 1.5 liter/min. (0.026 0.40 Gal/min)
|
Maximum Water Pressure
|
Integral pressure regulator rated 7 bar (101.5 PSI)
|
Cleaning
|
Ultrasonic cleaning
|
Calibration
|
Factory calibrated to 0 100 NTU, recalibration required every three months and after change of measuring range
|
Display
|
Multi-Line Liquid Crystal Backlit Display
|
Mounting
|
Wall mounting with 6 screws
|
Regulatory Compliance
|
Compliant to U.S. EPA 180.1 Compliant to DIN EN ISO 7027
|
Power
|
100–240 V AC, 47–63 Hz, 80 VA
|
Relays
|
Two Programmable, 120–240 V AC 2A Form C Relay
|
Analog Output
|
One powered 4–20 mA, 1000 W drive
|
Digital Output
|
Bi-directional RS-485, Modbus RTU
|
Environmental conditions
|
Storage Temperature Range: 1 °C ... 50 °C (34 °F 122 °F) Operating Temperature Range: 1 °C ... 50 °C (34 °F 122 °F) Relative Humidity: Up to 95 % (non-condensing) Not recommended for outdoor use. Altitude up to 2000 meters.
|
Certification
|
CE, cETLus CE Approved, LC listed to UL 61010-1 3rd Edition: 2012 LC: Certified to CSA 22.2 No.61010-1-12: 2012
|
Mechanics
|
Wetted Materials: Nylon, Wetted Materials: Nylon, Borosilicate Glass, Silicone, Polypropylene, Stainless Steel, Viton, Acety Housing: ABS (Double Insulated, Pollution Degree 2, Overvoltage Category II)
|
Weight
|
2.8 kg (6.2 lbs.)
|
Warranty
|
2 years on defects in quality according to § 10 terms of conditions
|