Measuring method
|
Electrochemical
|
Appropriate Electrode
|
Reference electrode VARiON® Ref, Measuring electrode VARiON®Plus NO3, Compensation electrode VARiON®Plus Cl
|
Measuring range/Resolution
|
NO3-N: 1 ... 1000 mg/l / 1 mg/l; 0.1 ... 100.0 mg/l / 0.1 mg/l NO3-: 5 ... 4500 mg/l / 5 mg/l; 0.5 ... 450.0 mg/l / 0.5 mg/l
|
Compensation Range
|
Cl-: 0.1 ... 1,000 mg/l / 0.1 mg/l
|
Measuring Accuracy in laboratory standard solutions
|
± 5 % of measured value ± 0.2 mg/l in standard solutions
|
Calibration Procedures
|
Matrix adjustment against any reference value, 2-point-calibration possible with multiple standard solution
|
Working Life (typically)
|
Reference electrode: 18 months, measuring and compensation electrode: 18 months (in typical application - municipal sewage plants)
|
Temperature Measurement and Compensation
|
Integrated NTC thermistor, Range 32 °F … 104 °F (0 °C … +40 °C), Accuracy ±0.5 K, Resolution 0.1 K, t95 < 20 s
|
Pressure Resistance
|
Maximum 0.2 bar (incl. SACIQ sensor connection cable, with installed electrodes)
|
Ambient Conditions
|
Operating temperature: 32 °F … 104 °F (0 °C … +40 °C), storing temperature: 32 °F … 104 °F (0 °C … +40 °C)
|
Electrical connections
|
2-wire shield cable with quick fastener to sensor
|
Electromagnetic Compatibility
|
EN 61326, Class B, FCC Class A; Intended for indispensable operation
|
Certifications
|
CE
|
Mechanical
|
Sensor body: V4A stainless steel 1.4571 Protective cup: POM Temperature sensor: V4A stainless steel 1.4571 Protection rating: IP 68 (0.2 bar, with installed electrodes) Electrode connector: POM
|
Weight
|
Approx. 1.48 lb (670 g, without electrode, without sensor connection cable)
|
Warranty
|
NitraLyt®Plus 700 IQ: 2 years Electrodes: 1 year for defects of quality
|