Digital pH/ORP armature to be used with electrodes, connectable to the IQ Sensor Net
SensoLyt® 700 IQ – digital pH/ORP armature with integrated preamplifier and temperature sensor as well as lightning protection to be connected to IQ Sensor Net:
By storing calibration values within the sensor, a "pre-calibrated pH-measurement" can be generated. Due to our quick-lock, the sensor can be disconnected and – after laboratory calibration – re-connected at the site easily. Inconvenient calibrations in the field under adverse conditions can be completely eliminated if there is an IQ connection in the lab. A special switch also allows automatic monitoring for glass breakage. The required replacement of the SensoLyt® electrode is quick and uncomplicated.
* SW: Sensor as sea water model (with plastic arming (POM))
detailed technical data see data sheet D2.03
Model
SensoLyt® 700 IQ
SensoLyt® 700 IQ SW*
Order No.
109170
109171
Measuring method
Potentiometric
Measuring range
Resolution
0,01 pH 1 mV
Accuracy
Depends on calibration± 0,2 pH; ± 20 mV
Temp. measurement
Integrated NTC, -5 °C ... +60 °C
Ambient conditions
Operating temperature: 0 °C ... +60 °C
Electrical connections
2-wire shield cable with quick fastener to sensor
Weight (without cable)
Approx. 970 g
Approx. 1.800 g
Electrodes for the IQ SENSOR NET and for analog transmitters pH 298
Technical buffer solutions for WTW meters with pH functionality
Sensor and connection cables for the IQ SENSOR NET
Extensions and holders to mount 1 to 3 IQ sensors or analog process sensors
Flow-through Armatures and weld-in sockets made of PVC or stainless steel. Please order suitable sensor adapter(ADA-X) separately.
Adapters to install 40 mm sensors in flow-through aramutres and vessels
IQ SENSOR NET pH/ORP armature,Operating manual ba15324d08, German
IQ SENSOR NET pH/ORP armature,Operating manual ba15324e08, English
IQ SENSOR NET pH/ORP armature,Operating manual ba15324i08, Italian
IQ SENSOR NET pH/ORP armature,Operating manual ba15324pt09, Portuguese
IQ SENSOR NET pH/ORP armature,Operating manual ba15324no10, Norwegian
IQ SENSOR NET pH/ORP armature,Operating manual ba15324s09, Spanish
IQ SENSOR NET pH/ORP armature,Operating manual ba15324sv10, Swedish
IQ SENSOR NET pH/ORP armature,Operating manual ba15324pl, Polish
Product flyer, 2 pages, German, 999252D"pH- und Redox-Messung für jedes Wasser: DIGITAL, VERLÄSSLICH, BEQUEM ZU KALIBRIEREN"
Flyer, 4 pages, German, 999102D"IQ SENSOR NET – das System: Sicher - Modular - Flexibel"
Product flyer, 2 pages, English, 999252US"pH and ORP Measurement for all Waters: DIGITAL, RELIABLE, CONVENIENT CALIBRATION"
Product flyer, 2 pages, Polish, 999252PL"Pomiar pH i redoks dla każdej wody: CYFROWY, NIEZAWODNY, Z WYGODNĄ KALIBRACJĄ"
Flyer, 4 pages, English, 999102US"IQ SENSOR NET – the range: Secure - modular - flexible"
Flyer, 4 pages, Polish, 999102PL"System IQ SENSOR NET: bezpieczny. modułowy. elastyczny."
Product flyer, 2 pages, French, 999252FR"Mesure de pH et rédox pour tout type d’eau : NUMÉRIQUE, FIABLE, ETALONNAGE AISÉ"
Flyer, 4 pages, French, 999102FR"IQ SENSOR NET - La gamme : sûr, flexible, modulable."
Product flyer, 2 pages, Italian, 999252IT"Misurazione di pH e ORP di acque reflue: AFFIDABILE E COMODA, CALIBRAZIONE DIGITALE"
USING WTW SPECTRAL SENSORSApplication note
MIT WTW-SPEKTRALSENSORENApplication note
Application note
Data sheet D2.03, English, 999226USDigital IQ pH/ORP armatures SensoLyt®
Data sheet D2.03, German, 999226DDigitale pH-/Redox-Armaturen SensoLyt®
Data sheet D2.29, German, 999389DKorrosionsbeständige IQ-Sensoren (alle SW-Versionen im Überblick)
Data sheet D2.29, English, 999389USCorrosion-resistant IQ sensors (all SW versions at a glance)
Go to cart/checkoutContinue shopping