MIQ Module Coupling at Front
|
Combined mechanical and electrical connection for rapid docking and removal of the MIQ/TC 2020 3G Terminal/Controller (configurated as Terminal) and for docking additional modules
|
MIQ Module Coupling at Rear
|
Combined mechanical and electrical connection, for rapid coupling to MIQ modules, up to 3 modules as a stack mounted unit possible
|
Cable Feeds
|
4 screw cable glands M 16 x 1.5
|
Terminal Connections
|
Screw terminal strips Terminal area for solid conductors: 0.2 ... 4.0 mm2 Terminal area for flexible conductors: 0.2 ... 2.5 mm2 accessible by opening cover
|
IQ Sensor Net Terminal Connections
|
Terminal connections for the IQ Sensor Net are available on each module and can be used as required: - for connecting sensors - as an input/output or for looping through/branching of the IQ Sensor Net cable
|
Other Functions
|
Two LEDs, yellow and red, for monitoring the operating voltage of the IQ Sensor Net; IQ Sensor Net connection, Integrated local identity function; Integrated switchable terminal resistor (SN terminator)
|
Electric Supply
|
Directly via the IQ Sensor Net
|
Ambient Conditions
|
Operating temperature: -4 °F … 131 °F (-20 °C … +55 °C); Storage temperature: -13 °F … 149 °F (-25 °C … +65 °C)
|
Housing Material
|
PC – 20 % GF (polycarbonate with 20 % fiberglass)
|
Protection Rating
|
IP67
|
IP 66
|
corresponding to NEMA 4X (not for direct conduit connections). Conduits need to be connected with flexible adapters (CC-Box), respectively with adapters CC-PM..
|
Dimensions (W x H x D)
|
5.67 x 5.67 x 2.05 in. (144 x 144 x 52 mm)
|
Weight
|
Approx. 1.1 pounds (0.5 kg)
|
Certifications
|
ETL, cETL (conforms with relevant UL and Canadian standards), CE
|
Electromagnetic Compatibility
|
EN 61326-1, Class B; FCC Class A
|
Integrated Overvoltage Protection
|
According to EN 61326-1 enhanced overvoltage protection for the entire system, implemented in each component
|
Connection Medium Cable
|
IQ Sensor Net cable SNCIQ or SNCIQ/UG (underground cable with additional PVC coating): 2-wire with shield; 2 x 0.75 mm2; Filler cord for easy connection of shield: 0.75 mm2; pressure resistant to 10 bar
|
Connection Characteristics
|
Energy and data transfer via 2 wire technique; resistant to reversed polarity; Comprehensive EMC shield control; cable topology within IQ Sensor Net system as required, e.g. in the form of a line, tree, star, multiple star; Total cable length: max. 1000 m/1094 yds (without signal amplifying), with signal amplifying module MIQ/JBR additional 1000 m/1094 yds (max 3000 m/3282 yds)
|
Warranty
|
3 years for defects of quality
|