UV spectral sensors for nitrite, nitrate, COD and BOD - WTW

UV spectral sensors for nitrite, nitrate, COD and BOD - WTW

  • SKU: 481055
Request a Quote
  • Description
    Digital, optical sensor for the IQ SENSOR NET
    NiCaVis® 700 IQ NI: UV probes with integrated ultrasonic cleaning for the reagent-free measurement of nitrate, nitrite and carbon parameters directly in the process:
    •Measurement directly in the process
    •No waiting for response times
    •Maintenance-free WTW ultrasonic cleaning
    •No wear parts, like wipers
    •Now available for surface water
    The full flexibility for your measurement:
    The high resolution of these sensors allows the precise distinction and measurement of nitrate and nitrite. At the same time, up to 3 carbon parameter such as COD, BOD, DOC, TOC, UVT or SAC are determined. This allows the complete monitoring and optimization of processes in the aeration of wastewater treatment plants.
  • Specifications

    Model

    NiCaVis® 701 IQ NI

    NiCaVis® 705 IQ NI

    Order No.

    481054

    481055

    * The UVT-254 value is standardized to 10 mm gap width.

    Measuring method

    Spectral Measurement in the UV Range (200 – 390 nm)

    Measuring gap
    (optical layer thickness)

    1 mm

    5 mm

    Application (optimized for)

    Municipal wastewater:

    Municipal wastewater:

    Measuring range and Resolution

    Inlet:

    NO3

    0.0 ... 300.0 mg/l 0.1 mg/l

    NO3-N

    0.00 ... 60.00 mg/l 0.01 mg/l

    NO2

    0.0 ... 120.0 mg/l 0.1 mg/l

    NO2-N

    0.00 ... 30.00 mg/l 0.01 mg/l

    COD total

    0 ... 20,000 mg/l 1 mg/l

    COD dissolv

    0 ... 12,500 mg/l 1 mg/l

    TOC

    0 ... 20,000 mg/l 1 mg/l

    DOC

    0 ... 12,500 mg/l 1 mg/l

    BOD

    0 ... 8,000 mg/l 1 mg/l

    SAC254 total

    0 ... 5,000 1/m 1 1/m

    UVT254 total*

    0 ... 100.0 % 0.1 %

    Aeration:

    NO3

    0.0 ... 300.0 mg/l 0.1 mg/l

    NO3-N

    0.00 ... 60.00 mg/l 0.01 mg/l

    NO2

    0.0 ... 120.0 mg/l 0.1 mg/l

    NO2-N

    0.00 ... 30.00 mg/l 0.01 mg/l

    COD dissolv

    0 ... 12,500 mg/l 1 mg/l

    DOC

    0 ... 12,500 mg/l 1 mg/l

    SAC254 total

    0 ... 5,000 1/m 1 1/m

    UVT254 total*

    0 ... 100.0 % 0.1 %

    Effluent:

    Effluent:

    NO3

    0.0 ... 750.0 mg/l 0.1 mg/l

    0.0 ... 250.0 mg/l 0.1 mg/l

    NO3-N

    0.0 ... 150.0 mg/l 0.1 mg/l

    0.00 ... 50.00 mg/l 0.01 mg/l

    NO2

    0.0 ... 300.0 mg/l 0.1 mg/l

    0.0 ... 100.0 mg/l 0.1 mg/l

    NO2-N

    0.00 ... 75.00 mg/l 0.01 mg/l

    0.00 ... 25.00 mg/l 0.01 mg/l

    COD total

    0 ... 4.000 mg/l 1 mg/l

    0.0 ... 800.0 mg/l 1 mg/l

    COD dissolv

    0 ... 4.000 mg/l 1 mg/l

    0.0 ... 800.0 mg/l 1 mg/l

    TOC

    0 ... 2.500 mg/l 1 mg/l

    0.0 ... 500.0 mg/l 1 mg/l

    DOC

    0 ... 2.500 mg/l 1 mg/l

    0.0 ... 500.0 mg/l 1 mg/l

    BOD

    0 ... 2.500 mg/l 1 mg/l

    0.0 ... 500.0 mg/l 1 mg/l

    SAC254 total

    0 ... 3.000 1/m 1 1/m

    0.0 ... 600.0 1/m 1 1/m

    UVT254 total*

    0 ... 100.0 % 0.1 %

    0.0 ... 100.0 % 0.1 %

    Flow rate

    ≤ 3 m/s

    Pressure Resistance

    Maximum 1 bar (incl. sensor connection cable)

    Electrical connections

    2-wire shield cable with quick fastener to sensor

    Electromagnetic Compatibility

    EN 61326. Class B. FCC Class A
    Intended for indispensable operation

    Certifications

    CE

    Mechanical

    Housing: Titan Grade 2. PEEK, Window: Sapphire glass
    Protection class: IP 68

    Weight (without cable)

    Approx. 8.82 lb (4 kg)

    Warranty

    2 years for defects in quality

Downloads