Thermoreactor CR 3200 - WTW

Thermoreactor CR 3200 - WTW

  • SKU: 1P22-
Request a Quote
  • Description
    Digestions of 24 round cells with pre-programmed and user-defined programs
    In total 24 round cells (16 mm) in two seperate heating blocks using the same program
    • 8 stored programs for easy and fast operation including COD fast digestion
    • 8 user-defined programs from room temperature/25-170°C with high temperature digestion
    • AQA/IQC with optional external temperature sensor for documentation

    Programs for standard parameters COD including fast digestion (148°C, 20 min.), total nitrogen, total phosphorus, TOC and heavy metals of galvanizing with automatic heating and switch-off quick. User defined programs can be easily entered and stored from room temperature respectively 25-170°C to perform special digestions and thus allow high temperature digestion.
    The very same program can be run for all vials at the same time; to perform two parallel digestion programs, the CR 4200 is the suitable choice. For the analytical quality assurance (AQA), the external temperature sensor TFK CR can be connected to the PC/printer instead of a round cell.

    Programs for standard parameters
    •1: 148 °C for 120 minutes
    •2: 120 °C for 30 minutes
    •3: 120 °C for 60 minutes
    •4: 120 °C for 120 minutes
    •5: 100 °C for 60 minutes
    •6: 148 °C for 20 minutes
    •7: 150 °C for 120 minutes
    •8: 100 °C for 30 minutes
  • Specifications
    Reactor type Dry temperature control device with safety cover
    Cell shafts 2 x 12 cell shafts for reaction cells, diameter: 16 ± 0.2 mm
    Reaction time setting program settings or user-defined programs up to 180 min
    Temperature setting stored program, user-defined settings from room temperature up to 170°C
    Controlling accuracy ± 1 °C ± 1 digit
    Temperature stability ± 0.5 K
    Overtemperature protection 190 °C ± 5 °C
    Dimensions D x W x H: 312 x 255 x 185 mm
    Weight: 4 kg
    Output unidirectional RS232 interface with 2 banana sockets for external temperature sensor (article-# 250 100), printer, PC
    Power supply 115/230 VAC 50/60 Hz switchable via fuse setting
    allowed voltage tolerance: ±15 % maximum power consumption: 560 W, Fuse 2x6.3 AT
    Protection Class IP 20 according to EN 60529
    Ambient temperature Operation +5 °C to +40 °C
    Climatic class 2 according to VDI/VDE 3540
      Relative humidity: Yearly mean: < 75 %
      30 days /year: 95 % Other days: 85 % Light dew: yes
      Site altitude: max. 5000 m above sea level
    EMC EN 61326-1
      FCC 47 CFR Part 15
    Test certificates cETLus, CE
    Safety standards EN 61010-1
      EN 61010-2-010
      UL 61010-1
      CAN/CSA C22.2#61010-1
      CAN/CSA C22.2#61010-2-010

Downloads

Related Products